Section 1

Preview this deck

Things to do when you xray a baby

Front

Star 0%
Star 0%
Star 0%
Star 0%
Star 0%

0.0

0 reviews

5
0
4
0
3
0
2
0
1
0

Active users

1

All-time users

1

Favorites

0

Last updated

1 year ago

Date created

Mar 1, 2020

Cards (314)

Section 1

(50 cards)

Things to do when you xray a baby

Front

Lower the kVP (you don't need as much to penetrate), keep mAs similar or slightly lower Dont use a grid

Back

Focal spot

Front

Part of the anode where bombardment/xray production takes place

Back

What does the II do when you mag?

Front

Simply not "minifying" as much. But it still needs the same amount of light at the end so it uses more mA. Double mag = double skin dose. (but kerma area dose stays the same bc smaller area)

Back

Change in unsharpness with increase in focal spot size

Front

Bigger focal spot = less sharp. Small focal spot = more sharp

Back

Flat panel detector components

Front

CsI input phosphor (convert photons to light) Photodiode (convert light to charge) Readout elements - read out charge directly (vs II which needs to convert back to light and then use TV display)

Back

Compton scatter

Front

Predominates at higher energies relative to coherent and photoelectric. Ionizes an outer shell electron and redirection of the xray (which still has energy) to somewhere random. This is the worst!

Back

Heel effect

Front

Xrays that are closer to the anode side had to travel through more attenuating material and are less intense. Remember "cathode to chest wall"

Back

Which focal spot determines the blur?

Front

The apparent focal spot (on the patient), not the actual focal spot on the anode

Back

Steeper (smaller) anode angle

Front

Makes a smaller apparent focal spot, but the actual focal spot is also smaller so heat dissipation can be a problem

Back

Average energy relative to kVP

Front

Approximately 50% (prob slightly less) of the kVP

Back

Types of mammo target/filter combos

Front

Moly/moly, moly/rho, rho/rho. No such thing as rho/moly

Back

Double the mA

Front

Double the intensity of the energy spectrum

Back

Which causes more dose increase, geometric mag or electronic mag?

Front

Geometric (Physical distance) mag (squared vs doubled)

Back

Relationship between Z and amount of Bremsstrahlung

Front

Directly proportional

Back

Auger electron

Front

Inner shell electron gets ionized, another drops down, emits energy but this energy ionizes another (outer shell) electron. No x-rays produced

Back

Material for mammo exit window

Front

Beryllium (vs pyrex glass for standard radiography) bc you dont want to attenuate these low energy xrays

Back

Max energy of the xray that is created at the anode

Front

Can't be higher than the energy of the electron that created it (which maxes at the voltage between the cathode and anode in kV)

Back

Heel "cutoff"

Front

Smaller (steeper) anode angle results in worse heel effect

Back

Focal spot in mammo

Front

Has to be really small! ~0.3mm. This does not tolerate heat as well. You have to lower the mA or you'll melt the anode. Also means longer exposure time

Back

Best energy of the beam to visualized iodinated contrast

Front

Just above the K-edge of iodine, so approx 34-35 keV

Back

K-shell binding energy proportional to...

Front

Z^2

Back

Soft tissue half value layer

Front

~3cm

Back

K edge of Iodine

Front

33.2 keV

Back

Linear attenuation coefficient

Front

The portion of photons that gets attenuated per unit thickness. Ex: LAC of 0.1/cm would mean that 10% of the beam gets attenuated per cm of material. Kind of inverse to half value layer.

Back

When do you never use a grid?

Front

Anything less than 10cm basically (either a baby or an extremity)

Back

Why is rhenium mixed with tungsten in a filament?

Front

Prevent cracking with repeated heating/cooling

Back

Actual focal spot vs apparent focal spot

Front

Actual focal spot is on the target. Apparent focal spot is on the patient

Back

Average energy of the xrays is based on the...

Front

kVP, unrelated to mA

Back

Dose change when you bring patient closer to the source

Front

Inverse Squared!

Back

How to improve heel effect

Front

Smaller film size, larger source to image distance, larger (less steep) anode angle

Back

Where do you want to stand as the operator?

Front

On the side of the receptor. This avoids the majority of the scatter

Back

Direct DR system (Selenium)

Front

No light involved (hence direct). Photon hits detector and gets turned into charge by amorphous selenium

Back

How does Z affect Compton scatter?

Front

Doesnt matter. Compton scatter directly proportional to tissue DENSITY

Back

Which new target/filter setup is for denser breasts?

Front

Rho/Al, or even Tungsten/Rho or Tungsten/Silver

Back

Smaller focal spot

Front

Better spatial resolution, worse heat dissipation

Back

Components of fluoro image intensifier (II)

Front

Input phosphor (Cs) turns xrays into light Photo-cathode turns light into charge Charge accelerated, hits output phosphor Output phosphor is smaller (concentrates) and turns back into light (minification gain)

Back

Fluoro focal spot vs general radiograph

Front

Fluoro focal spot is smaller (want less blurring)

Back

K edge of barium

Front

37.4 keV

Back

K shell binding energy of tungsten

Front

69.5 keV

Back

Breast tissue half value layer

Front

~1cm

Back

What elements tend to make Auger electrons?

Front

Lower Z. This is also why you pick high Z for anodes

Back

Which classic mammo target/filter setup is for denser breasts?

Front

Rho/rho

Back

How does Z affect photoelectric effect?

Front

Probability is directly associated with Z^3. So much more photoelectric effect at high Z

Back

What is classical/coherent (Rayleigh) scatter?

Front

Occurs in low energy situations (think mammo) usually <10keV. No ionization occurs. Can result in some loss of image contrast.

Back

"Blooming" of the focal spot

Front

Happens when you have high mA, low kVP. Imagine crowded low energy electrons that are bumping each other when theyre going from cathode to anode.

Back

Breast mag views

Front

No grid, use air gap. Even smaller focal spot (0.1mm) Even lower mA / longer exposure Smaller paddle

Back

Minimum energy (in eV) needed to be ionizing

Front

15eV

Back

Why compress the breast?

Front

Decrease scatter (improved contrast), lower kvP, less mA needed, less motion.

Back

Increase kVP by 15%

Front

Double intensity of energy spectrum

Back

Type of shielding to use when blocking beta emitters (like a syringe full of Y90)

Front

Low Z shield (like plastic) because high Z shield will create tons of Bremsstrahlung xrays

Back

Section 2

(50 cards)

What increases axial resolution?

Front

Higher frequency probe

Back

Dose limit for an IR fellow

Front

50mSv per year. In reality you get about 5mSv

Back

Bowtie filter

Front

Filter attenuates less in the center than in the edges so that you get a more homogeneous beam.

Back

Absorbed dose vs equivalent dose

Front

Equivalent dose takes into account the type of radiation (Ex: alpha particles are a **ton worse than xrays)

Back

When to use a narrow window

Front

When the things you're trying to tell apart are close in density (like stroke windows for white/gray matter)

Back

Compound imaging

Front

"Peeking around the corner" Cleaner edges Lose posterior shadowing

Back

What is the reconstruction kernel?

Front

Post-processing, tradeoff between spatial resolution and noise. Sharp kernel has high spatial resolution and tons of noise (think bone). Soft kernal has low spatial resolution and low noise. Dose is the SAME b/c its post-processing

Back

What is weighted CTDI?

Front

1/3 the central CTDI + 2/3 the peripheral CTDI

Back

What is DLP?

Front

Dose length product (DLP) = Volume CTDI * Scan length

Back

Dark streak artifact

Front

Look for arms down, streaks in the mediastinum. This is a kind of beam hardening artifact. Easiest fix is to tilt the gantry or position the patient with arms up.

Back

Frequency vs spatial pulse length

Front

The higher the frequency the smaller the spatial pulse length, the better your axial resolution

Back

Pitch > 1

Front

There's a gap between slices. Less dose. More spatial resolution

Back

Output power vs gain

Front

Output power = how big the wave is that you're sending out. Receiver gain is how much you tune up the strength of the pulse you get back.

Back

CT minimum slice thickness determined by

Front

Detector element aperture width (smaller is thinner)

Back

Pitch < 1

Front

There is overlap. More dose. Improved spatial resolution.

Back

Absorbed dose units

Front

In Gy (1 Joules/kg)

Back

Which has better SNR, the smooth kernel or the sharp kernel?

Front

The smooth kernel has better SNR, poorer spatial resolution.

Back

When to use wider window

Front

When the things you're trying to tell apart are wide apart in HU (like looking at a cancer in lung window)

Back

Ring artifact

Front

Defective detector element

Back

Output power vs lateral resolution

Front

Lower output power = better lateral resolution

Back

Iterative reconstruction

Front

Basically you can lower the dose and allegedly maintain imaging quality with advanced reconstruction

Back

How to increase SNR

Front

Thicker slices (fatter transmit bandwidth), larger FOV, smaller matrix, stronger magnet, narrower readout badwidth, biger voxel

Back

Effective dose units

Front

In Sieverts

Back

What is the effective dose?

Front

You need the DLP times a constant which depends on the body part

Back

What is axial resolution in US?

Front

Resolution in the direction of the probe (depth)

Back

What is a projection?

Front

All the rays in a given angle of the tube. Ray is a single beam to a single detector.

Back

Harmonics

Front

Transmit at one frequency and receive in another. Improves LATERAL resolution. Needs to travel a certain distance before you get this.

Back

Locations of collimation in CT

Front

Pre-patient (reduces dose on the source side) Post-patient (gets rid of scatter from patient on the detector side)

Back

What is volume CTDI?

Front

Weighted CTDI / pitch

Back

Relationship between HU and xray attenuation

Front

Increase HU by 10 = increase xray attenuation by 1%

Back

Pixel/voxel size relative to FOV and matrix

Front

Pixel size = Field of view / Matrix

Back

Is KAP related to source distance?

Front

No, independent. Entrance air kerma goes down with distance, but the area hit goes up.

Back

What is CTDI?

Front

Radiation dose that is normalized to the width of the beam. Based on a phantom

Back

Increased beam width effect on dose

Front

Does NOT change. mAs unchanged. Lower scan time but bigger chunk of tissue exposed

Back

Equivalent dose units

Front

In Sieverts

Back

Stair step artifact

Front

Reconstructions with slices that are too thick (non overlapping intervals)

Back

Where is lateral resolution the best?

Front

At the focal zone

Back

Kerma area product / dose area product

Front

Entrance air kerma * area of entrance - measured in Gy-cm^2

Back

Matrix size in CT

Front

512 x 512

Back

Equivalent dose vs effective dose

Front

Effective dose takes into account tissue weighting factor (how sensitive an organ is to radiation)

Back

CT mA compared to xray

Front

CT mA is higher (up to 1000) compared to 100-800mA for radiograph

Back

What is power doppler

Front

Just detection of presence of doppler shift and how much. Doesnt give directional information.

Back

Pitch

Front

Pitch = table movement per tube / beam width

Back

Doubling the xrays does what to the signal? What to the noise?

Front

Doubling xrays doubles the signal, increases noise by sqrt(2). So overall SNR improves by 1.4x

Back

Cupping artifact

Front

Center of the image looking darker, didnt get enough juice relative to the outside

Back

Incomplete projection artifact

Front

Guy is too fat, tissue hanging out outside of field of view causing computer to spaz. Reposition patient if possible

Back

Best angle for doppler

Front

Theoretically, 0 is best and 90 is worst (cosine) but you MUST have <60.

Back

Color vs power doppler

Front

Power is more sensitive No aliasing in power Power not affected by doppler angle Both can get FLASH artifact

Back

What is mechanical index in US?

Front

Indication of US beam's ability to cause cavitation/micromechanical damage

Back

Will mag increase the KAP?

Front

No. The entrance air kerma will be bigger but the area will be smaller

Back

Section 3

(50 cards)

What are stochastic effects?

Front

No threshold. As dose increases, probability of occurrence increases but the severity does not. Think about cancer risk with radiation

Back

Effects of acute radiation syndromes are seen above what cutoff?

Front

1Gy

Back

Dose limit for a fetus of an occupational worker

Front

500 mrem (5mGy) through the term of the pregnancy

Back

Dose calibrator mnemonic

Front

CLAG for daily, quarterly, annually, whenever moved

Back

CDTIvol phantoms

Front

Estimated on 16cm or 32cm phantom. If the patient is smaller than the phantom, dose is underestimated. If patient is larger, dose is overestimated.

Back

What is the max whole body annual dose allowable?

Front

50mSv / 5rem / 5000mrem

Back

Flow related enhancement is most commonly seen on..

Front

T1 weighted sequences at the end (entry) slice

Back

Threshold for permanent hair loss

Front

7Gy

Back

Units of exposure

Front

Roentgen (Coulomb/kg)

Back

FDA limits of mechanical index

Front

1.0 for baby ~2.0 for adult

Back

Threshold for main erythema reaction

Front

Main erythema rxn occurs ~10 days post exposure (after transient) and threshold is ~5Gy

Back

Threshold for skin exposure at which additional care and followup is instituted

Front

15Gy to single exposure field

Back

What does NRC regulation 10 CFR Part 35 deal with?

Front

Medical use of by-product radioactive material

Back

Annual dose limit to the public

Front

1mSv / 100mrem

Back

Max whole body annual dose for a pregnant worker

Front

5mSv / 500mrem

Back

Purpose of normalization scan in PET QA/QC

Front

Correct for nonuniformity of detector elements to a uniform source (varying efficiency along different lines of response)

Back

Thermal index and OB

Front

Avoid using color doppler on a fetus bc this requires higher power and increases thermal index

Back

Beta plus decay

Front

Occurs in proton rich environments. Proton turns into a neutron, emits a positron. Atomic number decreases

Back

Allowed difference from prescribed nucs dose by licensee

Front

Up to 20%

Back

Who defined the thresholds for deterministic effects?

Front

International Commission on Radiological Protection (ICRP)

Back

Ring down artifact

Front

Gas bubbles with water inside that resonates and keeps generating more return waves

Back

Who needs to wear radiation badges

Front

People who have the possibility of getting more than 10% annual dose limit (so more than 5mGy or 500mrem)

Back

Threshold for late erythema reaction

Front

~15Gy. This happens 8-10 weeks after exposure and can appear bluish 2/2 ischemia.

Back

Mirror image artifact

Front

Type of reverberation artifact with an angled reflector

Back

What is most adjusted by automatic exposure control?

Front

Exposure time to achieve appropriate exposure of the detector (ends up changing mAs)

Back

Thermal index in ultrasound

Front

Ability to cause heat. Thermal index of 1 raises the temperature of tissue 1 degree celcius. This is tissue specific

Back

Threshold of I-131 that needs written AU directive

Front

>30uCi (1.11Mbq). Or any therapeutic dosage of unsealed byproduct

Back

Anisotropy artifact

Front

Tendon looks hyperechoic when perpendicular and hypoechoic when at an angle (can simulate tear). Fix = keep probe steady at perpendicular

Back

Refraction artifact

Front

Happens related to interfaces.

Back

Threshold for permanent sterility in males from acute exposure

Front

6Gy

Back

Maximum MQSA glandular dose from single view

Front

3mGy per image on the phantom. No specific limits on an actual breast.

Back

What improves susceptibility artifact?

Front

Lower field strength, increasing receive/transmit bandwidth, spin echo (as opposed to GRE or EPI)

Back

Interval for dose calibrator accuracy test

Front

Annually

Back

What does 10 CFR 20 deal with?

Front

Dose exposure limits for workers

Back

Relative biological effectiveness

Front

Ratio of doses from diff types of radiation (xrays, alpha, beta) to cause a given effect.

Back

What is MQSA?

Front

Mammo Quality Standards Act Regulations

Back

Minimum distance needed for axial resolution in US

Front

Half of an SPL (spatial pulse length)

Back

Color bleed artifact

Front

Will look like color extending beyond vessel wall. Improve by turning down color gain

Back

Interval for dose calibrator linearity test

Front

Quarterly

Back

Twinkle artifact

Front

Looks like theres movement/flow when there isnt. bc rough surface causes split into complex wave pattern. Also could be related to "phase jitter"

Back

Acute dose threshold for permanent sterility

Front

6Gy in men, 3Gy in women.

Back

Threshold for cataracts from acute exposure

Front

0.5Gy. Cataracts will develop >20 years after acute exposure

Back

Threshold for temp hair loss

Front

2-5Gy

Back

Threshold for transient erythema

Front

2Gy / 200 rads

Back

Cancer incidence increase for radiation according to BEIR VII

Front

8% per 1Gy (0.08% per rad)

Back

What is multiple scan average dose (MSAD)?

Front

Avg dose to a slice in the central portion of a scan taking into account scatter from adjacent slices (compton)

Back

Max dose limitation for public from release of patients injected with radioactive materials

Front

5mSv / 500mrem

Back

Probe frequency vs mechanical index

Front

High frequency has low MI Low frequency has high MI

Back

Interval for dose calibrator constancy test

Front

Daily

Back

Aliasing ultrasound artifact

Front

Point where you screw up and lose data unless you increase sampling speed = Nyquist frequency. Appears as wrap-around Nyquist limit = 1/2 pulse repetition frequency

Back

Section 4

(50 cards)

Type of artifact seen with colloid inspisations

Front

Comet tail

Back

Things that cause comet tail artifact

Front

Cholesterol and colloid

Back

Dose threshold for risk of organ malformation to fetus

Front

100 mGy

Back

Dynamic range of digital vs plain film

Front

Digital has wider dynamic range

Back

In an II, what is flux gain?

Front

Acceleration of electrons towards the output phospor, increasing their energy

Back

Window width and contrast

Front

Wide window - decreased contrast Narrow window - increased contrast

Back

Annual MR QC

Front

Done by medical physicist, includes RF coil check, slice thickness and slice position accuracy

Back

Max dose to public accessible areas in 1 hour

Front

2mrem/hr (0.02mSv/hr)

Back

CR vs DR decentralized

Front

CR is centralized (C for centralized) DR is decentralized (D for decentralized)

Back

Diamagnetic materials

Front

Water and calcium, induced field opposes the external field

Back

Bucky factor

Front

mA required with grid / mA required without grid. Most common Bucky factor is 2-3

Back

Probability of photoelectric effect relative to photon energy

Front

INVERSELY proportional to the energy CUBED

Back

Resonance frequency per 1T

Front

~42 MHz

Back

Dynamic range curves for film vs digital

Front

Linear and wide for digital, curvilinear and narrow for film

Back

What changes to make to kVP / mAs in a kid

Front

Lower kVP (need less penetrating photons) and keep the mA same to slightly lower

Back

Comet tail artifact

Front

Looks kinda like ring down but quickly tapers off (ring down extends to deep image). It is a type of reverb artifact

Back

What is grid cut off?

Front

The grid blocks so many photons that you cause quantum mottle

Back

What is minification gain?

Front

Electrons from a large surface, concentrated on a smaller surface

Back

Linear vs mass attenuation

Front

Linear attenuation differs for water, vapor, and ice bc they take up different amounts of space Mass attenuation is the same regardless of phase because it's measured per unit mass

Back

Type of change to lower dose for a stone study

Front

Lower the mA and keep the kVP the same. You get more noise but this is ok bc youre looking for a super high contrast finding

Back

Pair production

Front

At super high keV and super high Z target (Colossus from xmen) the photon hits straight into the nucleus and emits a pair (1 electron and 1 positron). Positron then annihilates and you can image like PET

Back

Pixel pitch

Front

Spacing between pixels. The lower the pixel pitch the better the spatial resolution.

Back

Fluoro mA, exposure time, and focal spot relative to regular dx

Front

Fluoro has longer exposure times, lower mA (so you dont melt the anode) and a smaller focal spot

Back

Spatial resolution of DR vs CR

Front

DR is better, with direct better than indirect (no lateral dispersion)

Back

Threshold at which PE vs compton dominates

Front

PE dominates <30keV, compton dominates >30keV. Note that both types decrease with increasing energy, its just that there's more compton at those energies relatively

Back

Ideal energy for mammo

Front

Between 16-23keV, so you have to use a voltage of 25-30kVp (vs ~120 for chest radiograph)

Back

Tenth value layer (TVL)

Front

Thickness needed to attenuate 90% of the beam. Ends up being 3.xx HVLs

Back

When does an II need to be replaced?

Front

Conversion gain falls below 50%

Back

Indirect DR system

Front

Xrays hit CsI scintillator -> light -> converted to charge by photodiode -> readout by TFT (thin film transistor) array

Back

How long must calibration records of dose calibrators be kept?

Front

NRC says 3 years.

Back

Highest energy anode/filter pairs in mammo

Front

Tungsten/rho and Tungsten/silver

Back

Material in a CR cassette

Front

Barium fluorohalides

Back

Things that cause ring down artifact

Front

Trapped air bubbles in fluid

Back

Probability of photoelectric effect

Front

Directly proportional to Z^3 Inversely proportional to photon energy ^3

Back

Weekly MR QC

Front

Done by tech. Includes high/low contrast resolution, table positioning, center frequency

Back

Storage phosphor (CR) radiography

Front

Type of indirect, casette based system. Xray -> light -> charge

Back

Is there lateral dispersion in direct DR?

Front

No, no scintillator intermediate so no opportunity for light dispersion.

Back

SI unit for magnetism

Front

Tesla. 1T = 10,000 gauss. Earth field = 0.5 Gauss

Back

What does isobaric mean?

Front

The mass number didnt change. Beta +, beta -, and electron capture are all examples of isobaric transitions

Back

What happens to beam width, spatial resolution, and beam intensity at the focal zone

Front

At focal zone on US, beam width is narrowest, beam intensity highest, and spatial resolution highest

Back

Half life equation

Front

Half life = ln(2)/decay constant

Back

Spatial resolution values

Front

Screen film mammo: 15lp/mm Digital mammo: 7 lp/mm Digital radiopgraph: 3lp/mm CT: 0.7lp/mm MR: 0.3 lp/mm

Back

How does the dose of an II change as it gets older?

Front

Older II = more dose! Worse efficiency / conversion gain

Back

Filter pair thats never used in mammo

Front

Rho anode with a moly filter

Back

Paramagnetic materials

Front

Gado and deoxyhemoglobin, induced field enhances external field.

Back

Appearance of normal fetal lungs on US

Front

As echogenic as liver. If they look like fluid its effusions.

Back

Artifact associated with prospective cardiac gating

Front

Flash artifact

Back

Housing vs enclosure

Front

Housing surrounds everything, made of lead Enclosure is glass and surrounds the x-ray tube, maintaining a vacuum.

Back

Annual extremity (or any other organ but the eye) dose limit

Front

500mSV / 0.5Sv / 50 rem

Back

How many spot films per equivalent dose of 1minute fluoro?

Front

5-10

Back

Section 5

(50 cards)

Instrument used for measuring high doses

Front

Dose calibrator/ion chamber

Back

Things that increase spatial resolution of gamma camera collimator

Front

Thicker, longer septae. Narrow hole diameter

Back

NRC CFR part 19

Front

Notices, instructions, reports to workers

Back

Dose limit to the public per hour in unrestricted area

Front

No greater than 2mrem/hour in an unrestricted area

Back

What happens to image from diverging collimator

Front

Minifies a large object onto a smaller crystal

Back

Range for medium energy collimator

Front

200-400keV

Back

Criteria for "high radiation area"

Front

More than 1msV in 1 hour at 30cm

Back

Things that increase sensitivity of gamma camera collimator

Front

Thinner, shorter septae. Larger hole diameter

Back

How often does energy window need to be tested?

Front

Daily / before each different test using a different tracer.

Back

What happens to image from pinhole collimator

Front

Inverted and magnified

Back

Major spill of I131

Front

1mCi

Back

How often is intrinsic flood done?

Front

Weekly

Back

Where should a ring badge be worn?

Front

On the dominant hand, index finger, label facing inward toward source, under a glove

Back

Geiger counter dead time

Front

If you over load it (>100mR/h) it stops working until it dissipates.

Back

Number of disintegrations per Curie

Front

3.7 x 10^10

Back

Effective half life

Front

1 / effective = 1/physical + 1/biological

Back

How often is dose calibrator consistency checked?

Front

Daily

Back

Recommended counts for extrinsic and intrinsic flood

Front

5-10 million

Back

How often is extrinsic flood done?

Front

Once daily. This tests the collimators and the crystals

Back

Sensitivity/resolution of thick crystal

Front

Better sensitivity (chance of catching the gamma particle), worse spatial resolution (light produced moves before getting to PMT

Back

Problem with sodium iodine well counter

Front

Easily overwhelmed (if exceeds 5k counts/sec). Good for in vitro samples and wipe tests

Back

What is an isometric transition?

Front

Process of rearranging of electrons coupled to electron capture that causes gamma emission, good for imaging

Back

How often is dose calibrator accuracy checked?

Front

Annually

Back

Range for high energy collimator

Front

>400keV

Back

NRC CFR part 35

Front

Medical use of by-product material (human use of radioisotopes)

Back

Workhorse nucs collimator

Front

Parallel hole collimator

Back

Star artifact

Front

Caused by septal penetration of hexagonally oriented collimator holes by very high energy focal source (think post-therapy I131)

Back

Extrinsic flood

Front

WITH a collimator

Back

How long do you keep radioactive material?

Front

10 half lives

Back

Amount of nonuniformity that is allowable

Front

2-5% (1% in SPECT)

Back

Intrinsic flood

Front

WITHOUT a collimator

Back

How is sensitivity affected by distance for parallel hole collimator?

Front

NO change! Even though farther distance reduces counts by inverse square, it allows for greater field of view so no net change in counts.

Back

Relationship between sensitivity and spatial resolution of gamma camera collimator

Front

Inversely related

Back

What happens to image from converging hole collimator

Front

Magnifies WITHOUT inverting

Back

Nucs QC stuff that has to be done daily

Front

Extrinsic flood and energy window

Back

Mechanism of isotope production that is "carrier free"

Front

Cyclotron bombardment with charged particles -> transmutation

Back

Annual dose limit to the general public

Front

100mrem / 1mSv

Back

Impact of matrix size in gamma camera on acquisition time

Front

Larger matrix = longer acquisition time and reduced counts per pixel (worse SNR)

Back

Function of pulse height analyzer

Front

To discard background stuff that has energy too high/low to correspond to desired tracer

Back

How often does center of linearity need to be tested?

Front

Weekly

Back

How is gamma camera linearity and spatial resolution tested?

Front

By placing a lead bar phantom with parallel lines between a cobalt source and the gamma camera. Linearity is bad if the bars look wavy. Resolution is defined by ability to differentiate distinct bars

Back

How often does gamma camera image linearity and spatial resolution need to be tested?

Front

Weekly

Back

How often is dose calibrator linearity checked?

Front

Quarterly

Back

Nucs QC stuff that has to be done weekly

Front

Intrinsic flood, linearity and resolution, and center of rotation

Back

Major spill of In-111, Ga-67

Front

>10mCi

Back

How often is dose calibrator geometry checked

Front

Installation and anytime it's moved

Back

Major spill of Tc, Tl

Front

>100mCi

Back

Range for "low energy" collimator

Front

1-200keV

Back

NRC CFR part 20

Front

Radiation protection

Back

What does flood test for?

Front

Uniformity

Back

Section 6

(50 cards)

Thick dampening block (low q)

Front

"Thud" - more dampening, higher bandwidth, low Q, short spatial pulse length for better axial resolution

Back

Annual occupational lens dose limit

Front

20mSv / 2rem

Back

Situation in which there is a ton of reflection

Front

High impedance difference between adjacent tissues

Back

Total dose to embryo allowed in 9 month pregnancy

Front

5mSv / 500mrem

Back

Beam width artifact

Front

Signal from far zone falsely localizing into area of interest. Classically shown in bladder

Back

How soon do you have to report a "medical event"

Front

Call the doctor, patient, and NRC within 24 hours. Write them a letter within 15 days

Back

Side lobe artifact

Front

Artifact that happens when stuff from side lobes gets registered like it came from the main beam. Classic example = "pseudosludge" in the gallbladder or bladder

Back

Echogenicity vs scatter amplitude

Front

Hyperechoic = high scatter amplitude Hypoechoic = lower scatter amplitude

Back

Reverberation artifact

Front

Due to reflections between two parallel highly reflective surfaces. Looks like multiple equidistantly spaced linear reflections

Back

How often do you do a "blank scan" for PET?

Front

Daily. This is the PET equivalent of uniformity

Back

Loss of how many dB corresponds with 50% loss in signal intensity?

Front

-3dB = 50% loss in signal intensity

Back

Rarefaction

Front

The relaxed (low pressure) part of the sound wave (vs compression)

Back

Type of crystal used in PET

Front

BGO, LSO, or LYSO

Back

How to tell uncorrected from attenuation correction

Front

1. Skin is hot on uncorrected 2. Lung is hot on uncorrected

Back

Type of crystal used in planar imaging

Front

NaI

Back

Truncation artifact in PET

Front

When a fat dude cant fit in a CT and part of him gets left out of the image, but he can fit in the PET, and the attenuation correction is all jacked up as a result

Back

Radiochemical purity

Front

Testing for free Tc. Use thin layer chromatography.

Back

Fasting duration prior to PET

Front

At least 4 hours, longer to minimize cardiac activity

Back

Absorption in ultrasound

Front

Sound energy gets turned into heat. This increases with frequency.

Back

Thickness of crystal ~ frequency

Front

Thick crystal - low frequency Thin crystal - high frequency

Back

Definition of "half value thickness" in ultrasound

Front

Thickness of tissue that causes a reduction of ultrasound intensity by 3dB

Back

Criteria for "very high radiation area"

Front

More than 5 gray in 1 hour at 1m

Back

Distance an F18 positron travels in tissue

Front

~1mm

Back

Criteria for a "medical event"

Front

Wrong dose, wrong patient, wrong site (has to be off by over 20%) AND has to cause harm to the patient (whoe body dose >5rem /50mSv or single organ dose >50rem)

Back

What is impedance?

Front

Degree of "stiffness" of a tissue. Product of the velocity of sound in the medium and the density of the medium

Back

Narrow receive bandwidth results in...

Front

Better SNR, worse type 1 chemical shift, worse spatial resolution

Back

Unit of impedance

Front

Rayl

Back

Limit of how much the dose can be off from what you order

Front

20% via the NRC, 10% in some agreement states

Back

Steeper (bigger) receive bandwidth results in...

Front

Poorer SNR, better spatial resolution, better suppression of type 1 chemical shift

Back

Type of transducer that gets more side lobe artifact

Front

Linear array transducers

Back

Half life of Mo-99

Front

67 hours

Back

Minimum distance between objects in order to be resolved on axial resolution

Front

1/2 the spatial pulse length. This is smaller (better axial res) on low q / heavy damping

Back

When testing radionuclide purity, what do you assay for first?

Front

Mo first, to prevent issues with residual charge

Back

SUV estimation in fat people

Front

You'll overestimate SUV

Back

Thin damping (high Q)

Front

"Ding" - less dampening, narrow bandwidth, long spatial pulse length. good for doppler

Back

Chemical purity

Front

Testing for Al breakthrough. Test is with pH paper. Allowed amount is <10 microgram Al per 1mL

Back

Speed that US machine assumes sound travels at

Front

1540m/s

Back

What changes when ultrasound goes through different mediums? Wavelength or frequency?

Front

Speed of sound diff in diff media. As speed changes, frequency stays the same so wavelength has to change.

Back

Bucket setup imbalance

Front

Dark block rotating on sinogram

Back

How does frequency relate to scatter?

Front

Higher frequency = smaller wavelength = surfaces less smooth = more scatter

Back

Elevation resolution

Front

Resolution in the plane orthogonal to lateral resolution. Dependent on transducer element height

Back

Annual occupational total equivalent organ dose

Front

500mSv / 50rem. Same as extremity (hand) dose

Back

How often do you do a normalization scan for PET?

Front

Monthly. You scan a point source in the FOV

Back

Distribution of free Tc

Front

Salivary, thyroid, stomach

Back

What does a change of +/- 10dB have on the power

Front

+10dB = 10x power, -10db = 1/10th power

Back

Causes of free Tc

Front

Not enough tin, air in the syringe

Back

Radionuclide purity

Front

Testing for Mo breakthrough. You want less than 0.15MICROCi Mo per 1milCi of Tc - at the time of ADMINISTRATION

Back

Annual occupational whole body dose limit

Front

50mSv / 5rem

Back

Relationship between transducer frequency and near field length

Front

Higher frequency, longer near field

Back

How does axial resolution change with depth?

Front

It doesnt. Only lateral resolution does

Back

Section 7

(14 cards)

Annual MR scanner QC is done by the...

Front

Physicist

Back

Weekly MR scanner QC is done by the..

Front

Tech

Back

Crosstalk artifact

Front

Excitation into an adjacent slice, fix by reducing overlap (having a gap)

Back

Black blood cardiac MRI

Front

Double inversion recovery spin echo sequence

Back

FDA limit for specific absorption rate (SAR)

Front

4W/kg over 15 minutes

Back

Situation where neurostimulation occurs in MRI

Front

High gradient switching

Back

Dielectric effect is worse with...

Front

A stronger magnet

Back

Type of sequence in breast MRI for implant rupture

Front

Fat and water saturated (only silicone will be bright)

Back

Bright blood cardiac MRI

Front

Gradient sequences. SSFP is closer to a T2

Back

Delayed gad image sequence type cardiac MRI

Front

Inversion recovery. Null myocardium.

Back

Components of calculating SAR

Front

Magnet strength (squared) Flip angle (squared) Duty cycle (inversely related to TR and linear)

Back

Maximum MRI noise allowed by FDA

Front

140dB

Back

Eddy currents

Front

looks like distortion, most severe with DWI sequences

Back

Super long inversion time on cardiac imaging, to the point where blood is darker than myocardium.

Front

Hint that you're looking for amyloid

Back