Lower the kVP (you don't need as much to penetrate), keep mAs similar or slightly lower
Dont use a grid
Back
Focal spot
Front
Part of the anode where bombardment/xray production takes place
Back
What does the II do when you mag?
Front
Simply not "minifying" as much. But it still needs the same amount of light at the end so it uses more mA. Double mag = double skin dose. (but kerma area dose stays the same bc smaller area)
Back
Change in unsharpness with increase in focal spot size
Front
Bigger focal spot = less sharp. Small focal spot = more sharp
Back
Flat panel detector components
Front
CsI input phosphor (convert photons to light)
Photodiode (convert light to charge)
Readout elements - read out charge directly (vs II which needs to convert back to light and then use TV display)
Back
Compton scatter
Front
Predominates at higher energies relative to coherent and photoelectric. Ionizes an outer shell electron and redirection of the xray (which still has energy) to somewhere random. This is the worst!
Back
Heel effect
Front
Xrays that are closer to the anode side had to travel through more attenuating material and are less intense. Remember "cathode to chest wall"
Back
Which focal spot determines the blur?
Front
The apparent focal spot (on the patient), not the actual focal spot on the anode
Back
Steeper (smaller) anode angle
Front
Makes a smaller apparent focal spot, but the actual focal spot is also smaller so heat dissipation can be a problem
Back
Average energy relative to kVP
Front
Approximately 50% (prob slightly less) of the kVP
Back
Types of mammo target/filter combos
Front
Moly/moly, moly/rho, rho/rho. No such thing as rho/moly
Back
Double the mA
Front
Double the intensity of the energy spectrum
Back
Which causes more dose increase, geometric mag or electronic mag?
Front
Geometric (Physical distance) mag (squared vs doubled)
Back
Relationship between Z and amount of Bremsstrahlung
Front
Directly proportional
Back
Auger electron
Front
Inner shell electron gets ionized, another drops down, emits energy but this energy ionizes another (outer shell) electron. No x-rays produced
Back
Material for mammo exit window
Front
Beryllium (vs pyrex glass for standard radiography) bc you dont want to attenuate these low energy xrays
Back
Max energy of the xray that is created at the anode
Front
Can't be higher than the energy of the electron that created it (which maxes at the voltage between the cathode and anode in kV)
Back
Heel "cutoff"
Front
Smaller (steeper) anode angle results in worse heel effect
Back
Focal spot in mammo
Front
Has to be really small! ~0.3mm. This does not tolerate heat as well. You have to lower the mA or you'll melt the anode. Also means longer exposure time
Back
Best energy of the beam to visualized iodinated contrast
Front
Just above the K-edge of iodine, so approx 34-35 keV
Back
K-shell binding energy proportional to...
Front
Z^2
Back
Soft tissue half value layer
Front
~3cm
Back
K edge of Iodine
Front
33.2 keV
Back
Linear attenuation coefficient
Front
The portion of photons that gets attenuated per unit thickness. Ex: LAC of 0.1/cm would mean that 10% of the beam gets attenuated per cm of material. Kind of inverse to half value layer.
Back
When do you never use a grid?
Front
Anything less than 10cm basically (either a baby or an extremity)
Back
Why is rhenium mixed with tungsten in a filament?
Front
Prevent cracking with repeated heating/cooling
Back
Actual focal spot vs apparent focal spot
Front
Actual focal spot is on the target. Apparent focal spot is on the patient
Back
Average energy of the xrays is based on the...
Front
kVP, unrelated to mA
Back
Dose change when you bring patient closer to the source
Front
Inverse Squared!
Back
How to improve heel effect
Front
Smaller film size, larger source to image distance, larger (less steep) anode angle
Back
Where do you want to stand as the operator?
Front
On the side of the receptor. This avoids the majority of the scatter
Back
Direct DR system (Selenium)
Front
No light involved (hence direct). Photon hits detector and gets turned into charge by amorphous selenium
Back
How does Z affect Compton scatter?
Front
Doesnt matter. Compton scatter directly proportional to tissue DENSITY
Back
Which new target/filter setup is for denser breasts?
Front
Rho/Al, or even Tungsten/Rho or Tungsten/Silver
Back
Smaller focal spot
Front
Better spatial resolution, worse heat dissipation
Back
Components of fluoro image intensifier (II)
Front
Input phosphor (Cs) turns xrays into light
Photo-cathode turns light into charge
Charge accelerated, hits output phosphor
Output phosphor is smaller (concentrates) and turns back into light (minification gain)
Back
Fluoro focal spot vs general radiograph
Front
Fluoro focal spot is smaller (want less blurring)
Back
K edge of barium
Front
37.4 keV
Back
K shell binding energy of tungsten
Front
69.5 keV
Back
Breast tissue half value layer
Front
~1cm
Back
What elements tend to make Auger electrons?
Front
Lower Z. This is also why you pick high Z for anodes
Back
Which classic mammo target/filter setup is for denser breasts?
Front
Rho/rho
Back
How does Z affect photoelectric effect?
Front
Probability is directly associated with Z^3. So much more photoelectric effect at high Z
Back
What is classical/coherent (Rayleigh) scatter?
Front
Occurs in low energy situations (think mammo) usually <10keV. No ionization occurs. Can result in some loss of image contrast.
Back
"Blooming" of the focal spot
Front
Happens when you have high mA, low kVP. Imagine crowded low energy electrons that are bumping each other when theyre going from cathode to anode.
Back
Breast mag views
Front
No grid, use air gap.
Even smaller focal spot (0.1mm)
Even lower mA / longer exposure
Smaller paddle
Back
Minimum energy (in eV) needed to be ionizing
Front
15eV
Back
Why compress the breast?
Front
Decrease scatter (improved contrast), lower kvP, less mA needed, less motion.
Back
Increase kVP by 15%
Front
Double intensity of energy spectrum
Back
Type of shielding to use when blocking beta emitters (like a syringe full of Y90)
Front
Low Z shield (like plastic) because high Z shield will create tons of Bremsstrahlung xrays
Back
Section 2
(50 cards)
What increases axial resolution?
Front
Higher frequency probe
Back
Dose limit for an IR fellow
Front
50mSv per year. In reality you get about 5mSv
Back
Bowtie filter
Front
Filter attenuates less in the center than in the edges so that you get a more homogeneous beam.
Back
Absorbed dose vs equivalent dose
Front
Equivalent dose takes into account the type of radiation (Ex: alpha particles are a **ton worse than xrays)
Back
When to use a narrow window
Front
When the things you're trying to tell apart are close in density (like stroke windows for white/gray matter)
Back
Compound imaging
Front
"Peeking around the corner"
Cleaner edges
Lose posterior shadowing
Back
What is the reconstruction kernel?
Front
Post-processing, tradeoff between spatial resolution and noise. Sharp kernel has high spatial resolution and tons of noise (think bone). Soft kernal has low spatial resolution and low noise.
Dose is the SAME b/c its post-processing
Look for arms down, streaks in the mediastinum. This is a kind of beam hardening artifact. Easiest fix is to tilt the gantry or position the patient with arms up.
Back
Frequency vs spatial pulse length
Front
The higher the frequency the smaller the spatial pulse length, the better your axial resolution
Back
Pitch > 1
Front
There's a gap between slices. Less dose. More spatial resolution
Back
Output power vs gain
Front
Output power = how big the wave is that you're sending out. Receiver gain is how much you tune up the strength of the pulse you get back.
Back
CT minimum slice thickness determined by
Front
Detector element aperture width (smaller is thinner)
Back
Pitch < 1
Front
There is overlap. More dose. Improved spatial resolution.
Back
Absorbed dose units
Front
In Gy (1 Joules/kg)
Back
Which has better SNR, the smooth kernel or the sharp kernel?
Front
The smooth kernel has better SNR, poorer spatial resolution.
Back
When to use wider window
Front
When the things you're trying to tell apart are wide apart in HU (like looking at a cancer in lung window)
Back
Ring artifact
Front
Defective detector element
Back
Output power vs lateral resolution
Front
Lower output power = better lateral resolution
Back
Iterative reconstruction
Front
Basically you can lower the dose and allegedly maintain imaging quality with advanced reconstruction
You need the DLP times a constant which depends on the body part
Back
What is axial resolution in US?
Front
Resolution in the direction of the probe (depth)
Back
What is a projection?
Front
All the rays in a given angle of the tube. Ray is a single beam to a single detector.
Back
Harmonics
Front
Transmit at one frequency and receive in another. Improves LATERAL resolution. Needs to travel a certain distance before you get this.
Back
Locations of collimation in CT
Front
Pre-patient (reduces dose on the source side)
Post-patient (gets rid of scatter from patient on the detector side)
Back
What is volume CTDI?
Front
Weighted CTDI / pitch
Back
Relationship between HU and xray attenuation
Front
Increase HU by 10 = increase xray attenuation by 1%
Back
Pixel/voxel size relative to FOV and matrix
Front
Pixel size = Field of view / Matrix
Back
Is KAP related to source distance?
Front
No, independent. Entrance air kerma goes down with distance, but the area hit goes up.
Back
What is CTDI?
Front
Radiation dose that is normalized to the width of the beam. Based on a phantom
Back
Increased beam width effect on dose
Front
Does NOT change. mAs unchanged. Lower scan time but bigger chunk of tissue exposed
Back
Equivalent dose units
Front
In Sieverts
Back
Stair step artifact
Front
Reconstructions with slices that are too thick (non overlapping intervals)
Back
Where is lateral resolution the best?
Front
At the focal zone
Back
Kerma area product / dose area product
Front
Entrance air kerma * area of entrance - measured in Gy-cm^2
Back
Matrix size in CT
Front
512 x 512
Back
Equivalent dose vs effective dose
Front
Effective dose takes into account tissue weighting factor (how sensitive an organ is to radiation)
Back
CT mA compared to xray
Front
CT mA is higher (up to 1000) compared to 100-800mA for radiograph
Back
What is power doppler
Front
Just detection of presence of doppler shift and how much. Doesnt give directional information.
Back
Pitch
Front
Pitch = table movement per tube / beam width
Back
Doubling the xrays does what to the signal? What to the noise?
Front
Doubling xrays doubles the signal, increases noise by sqrt(2). So overall SNR improves by 1.4x
Back
Cupping artifact
Front
Center of the image looking darker, didnt get enough juice relative to the outside
Back
Incomplete projection artifact
Front
Guy is too fat, tissue hanging out outside of field of view causing computer to spaz. Reposition patient if possible
Back
Best angle for doppler
Front
Theoretically, 0 is best and 90 is worst (cosine) but you MUST have <60.
Back
Color vs power doppler
Front
Power is more sensitive
No aliasing in power
Power not affected by doppler angle
Both can get FLASH artifact
Back
What is mechanical index in US?
Front
Indication of US beam's ability to cause cavitation/micromechanical damage
Back
Will mag increase the KAP?
Front
No. The entrance air kerma will be bigger but the area will be smaller
Back
Section 3
(50 cards)
What are stochastic effects?
Front
No threshold. As dose increases, probability of occurrence increases but the severity does not. Think about cancer risk with radiation
Back
Effects of acute radiation syndromes are seen above what cutoff?
Front
1Gy
Back
Dose limit for a fetus of an occupational worker
Front
500 mrem (5mGy) through the term of the pregnancy
Back
Dose calibrator mnemonic
Front
CLAG for daily, quarterly, annually, whenever moved
Back
CDTIvol phantoms
Front
Estimated on 16cm or 32cm phantom. If the patient is smaller than the phantom, dose is underestimated. If patient is larger, dose is overestimated.
Back
What is the max whole body annual dose allowable?
Front
50mSv / 5rem / 5000mrem
Back
Flow related enhancement is most commonly seen on..
Front
T1 weighted sequences at the end (entry) slice
Back
Threshold for permanent hair loss
Front
7Gy
Back
Units of exposure
Front
Roentgen (Coulomb/kg)
Back
FDA limits of mechanical index
Front
1.0 for baby
~2.0 for adult
Back
Threshold for main erythema reaction
Front
Main erythema rxn occurs ~10 days post exposure (after transient) and threshold is ~5Gy
Back
Threshold for skin exposure at which additional care and followup is instituted
Front
15Gy to single exposure field
Back
What does NRC regulation 10 CFR Part 35 deal with?
Front
Medical use of by-product radioactive material
Back
Annual dose limit to the public
Front
1mSv / 100mrem
Back
Max whole body annual dose for a pregnant worker
Front
5mSv / 500mrem
Back
Purpose of normalization scan in PET QA/QC
Front
Correct for nonuniformity of detector elements to a uniform source (varying efficiency along different lines of response)
Back
Thermal index and OB
Front
Avoid using color doppler on a fetus bc this requires higher power and increases thermal index
Back
Beta plus decay
Front
Occurs in proton rich environments. Proton turns into a neutron, emits a positron. Atomic number decreases
Back
Allowed difference from prescribed nucs dose by licensee
Front
Up to 20%
Back
Who defined the thresholds for deterministic effects?
Front
International Commission on Radiological Protection (ICRP)
Back
Ring down artifact
Front
Gas bubbles with water inside that resonates and keeps generating more return waves
Back
Who needs to wear radiation badges
Front
People who have the possibility of getting more than 10% annual dose limit (so more than 5mGy or 500mrem)
Back
Threshold for late erythema reaction
Front
~15Gy. This happens 8-10 weeks after exposure and can appear bluish 2/2 ischemia.
Back
Mirror image artifact
Front
Type of reverberation artifact with an angled reflector
Back
What is most adjusted by automatic exposure control?
Front
Exposure time to achieve appropriate exposure of the detector (ends up changing mAs)
Back
Thermal index in ultrasound
Front
Ability to cause heat. Thermal index of 1 raises the temperature of tissue 1 degree celcius. This is tissue specific
Back
Threshold of I-131 that needs written AU directive
Front
>30uCi (1.11Mbq). Or any therapeutic dosage of unsealed byproduct
Back
Anisotropy artifact
Front
Tendon looks hyperechoic when perpendicular and hypoechoic when at an angle (can simulate tear). Fix = keep probe steady at perpendicular
Back
Refraction artifact
Front
Happens related to interfaces.
Back
Threshold for permanent sterility in males from acute exposure
Front
6Gy
Back
Maximum MQSA glandular dose from single view
Front
3mGy per image on the phantom. No specific limits on an actual breast.
Back
What improves susceptibility artifact?
Front
Lower field strength, increasing receive/transmit bandwidth, spin echo (as opposed to GRE or EPI)
Back
Interval for dose calibrator accuracy test
Front
Annually
Back
What does 10 CFR 20 deal with?
Front
Dose exposure limits for workers
Back
Relative biological effectiveness
Front
Ratio of doses from diff types of radiation (xrays, alpha, beta) to cause a given effect.
Back
What is MQSA?
Front
Mammo Quality Standards Act Regulations
Back
Minimum distance needed for axial resolution in US
Front
Half of an SPL (spatial pulse length)
Back
Color bleed artifact
Front
Will look like color extending beyond vessel wall. Improve by turning down color gain
Back
Interval for dose calibrator linearity test
Front
Quarterly
Back
Twinkle artifact
Front
Looks like theres movement/flow when there isnt. bc rough surface causes split into complex wave pattern. Also could be related to "phase jitter"
Back
Acute dose threshold for permanent sterility
Front
6Gy in men, 3Gy in women.
Back
Threshold for cataracts from acute exposure
Front
0.5Gy. Cataracts will develop >20 years after acute exposure
Back
Threshold for temp hair loss
Front
2-5Gy
Back
Threshold for transient erythema
Front
2Gy / 200 rads
Back
Cancer incidence increase for radiation according to BEIR VII
Front
8% per 1Gy (0.08% per rad)
Back
What is multiple scan average dose (MSAD)?
Front
Avg dose to a slice in the central portion of a scan taking into account scatter from adjacent slices (compton)
Back
Max dose limitation for public from release of patients injected with radioactive materials
Front
5mSv / 500mrem
Back
Probe frequency vs mechanical index
Front
High frequency has low MI
Low frequency has high MI
Back
Interval for dose calibrator constancy test
Front
Daily
Back
Aliasing ultrasound artifact
Front
Point where you screw up and lose data unless you increase sampling speed = Nyquist frequency.
Appears as wrap-around
Nyquist limit = 1/2 pulse repetition frequency
Back
Section 4
(50 cards)
Type of artifact seen with colloid inspisations
Front
Comet tail
Back
Things that cause comet tail artifact
Front
Cholesterol and colloid
Back
Dose threshold for risk of organ malformation to fetus
Front
100 mGy
Back
Dynamic range of digital vs plain film
Front
Digital has wider dynamic range
Back
In an II, what is flux gain?
Front
Acceleration of electrons towards the output phospor, increasing their energy
Done by medical physicist, includes RF coil check, slice thickness and slice position accuracy
Back
Max dose to public accessible areas in 1 hour
Front
2mrem/hr (0.02mSv/hr)
Back
CR vs DR decentralized
Front
CR is centralized (C for centralized)
DR is decentralized (D for decentralized)
Back
Diamagnetic materials
Front
Water and calcium, induced field opposes the external field
Back
Bucky factor
Front
mA required with grid / mA required without grid. Most common Bucky factor is 2-3
Back
Probability of photoelectric effect relative to photon energy
Front
INVERSELY proportional to the energy CUBED
Back
Resonance frequency per 1T
Front
~42 MHz
Back
Dynamic range curves for film vs digital
Front
Linear and wide for digital, curvilinear and narrow for film
Back
What changes to make to kVP / mAs in a kid
Front
Lower kVP (need less penetrating photons) and keep the mA same to slightly lower
Back
Comet tail artifact
Front
Looks kinda like ring down but quickly tapers off (ring down extends to deep image). It is a type of reverb artifact
Back
What is grid cut off?
Front
The grid blocks so many photons that you cause quantum mottle
Back
What is minification gain?
Front
Electrons from a large surface, concentrated on a smaller surface
Back
Linear vs mass attenuation
Front
Linear attenuation differs for water, vapor, and ice bc they take up different amounts of space
Mass attenuation is the same regardless of phase because it's measured per unit mass
Back
Type of change to lower dose for a stone study
Front
Lower the mA and keep the kVP the same. You get more noise but this is ok bc youre looking for a super high contrast finding
Back
Pair production
Front
At super high keV and super high Z target (Colossus from xmen) the photon hits straight into the nucleus and emits a pair (1 electron and 1 positron). Positron then annihilates and you can image like PET
Back
Pixel pitch
Front
Spacing between pixels. The lower the pixel pitch the better the spatial resolution.
Back
Fluoro mA, exposure time, and focal spot relative to regular dx
Front
Fluoro has longer exposure times, lower mA (so you dont melt the anode) and a smaller focal spot
Back
Spatial resolution of DR vs CR
Front
DR is better, with direct better than indirect (no lateral dispersion)
Back
Threshold at which PE vs compton dominates
Front
PE dominates <30keV, compton dominates >30keV. Note that both types decrease with increasing energy, its just that there's more compton at those energies relatively
Back
Ideal energy for mammo
Front
Between 16-23keV, so you have to use a voltage of 25-30kVp (vs ~120 for chest radiograph)
Back
Tenth value layer (TVL)
Front
Thickness needed to attenuate 90% of the beam. Ends up being 3.xx HVLs
Back
When does an II need to be replaced?
Front
Conversion gain falls below 50%
Back
Indirect DR system
Front
Xrays hit CsI scintillator -> light -> converted to charge by photodiode -> readout by TFT (thin film transistor) array
Back
How long must calibration records of dose calibrators be kept?
Front
NRC says 3 years.
Back
Highest energy anode/filter pairs in mammo
Front
Tungsten/rho and Tungsten/silver
Back
Material in a CR cassette
Front
Barium fluorohalides
Back
Things that cause ring down artifact
Front
Trapped air bubbles in fluid
Back
Probability of photoelectric effect
Front
Directly proportional to Z^3
Inversely proportional to photon energy ^3
Back
Weekly MR QC
Front
Done by tech. Includes high/low contrast resolution, table positioning, center frequency
Back
Storage phosphor (CR) radiography
Front
Type of indirect, casette based system. Xray -> light -> charge
Back
Is there lateral dispersion in direct DR?
Front
No, no scintillator intermediate so no opportunity for light dispersion.
Back
SI unit for magnetism
Front
Tesla. 1T = 10,000 gauss. Earth field = 0.5 Gauss
Back
What does isobaric mean?
Front
The mass number didnt change. Beta +, beta -, and electron capture are all examples of isobaric transitions
Back
What happens to beam width, spatial resolution, and beam intensity at the focal zone
Front
At focal zone on US, beam width is narrowest, beam intensity highest, and spatial resolution highest
Back
Half life equation
Front
Half life = ln(2)/decay constant
Back
Spatial resolution values
Front
Screen film mammo: 15lp/mm
Digital mammo: 7 lp/mm
Digital radiopgraph: 3lp/mm
CT: 0.7lp/mm
MR: 0.3 lp/mm
Back
How does the dose of an II change as it gets older?
Front
Older II = more dose! Worse efficiency / conversion gain
Back
Filter pair thats never used in mammo
Front
Rho anode with a moly filter
Back
Paramagnetic materials
Front
Gado and deoxyhemoglobin, induced field enhances external field.
Back
Appearance of normal fetal lungs on US
Front
As echogenic as liver. If they look like fluid its effusions.
Back
Artifact associated with prospective cardiac gating
Front
Flash artifact
Back
Housing vs enclosure
Front
Housing surrounds everything, made of lead
Enclosure is glass and surrounds the x-ray tube, maintaining a vacuum.
Back
Annual extremity (or any other organ but the eye) dose limit
Front
500mSV / 0.5Sv / 50 rem
Back
How many spot films per equivalent dose of 1minute fluoro?
Front
5-10
Back
Section 5
(50 cards)
Instrument used for measuring high doses
Front
Dose calibrator/ion chamber
Back
Things that increase spatial resolution of gamma camera collimator
Front
Thicker, longer septae. Narrow hole diameter
Back
NRC CFR part 19
Front
Notices, instructions, reports to workers
Back
Dose limit to the public per hour in unrestricted area
Front
No greater than 2mrem/hour in an unrestricted area
Back
What happens to image from diverging collimator
Front
Minifies a large object onto a smaller crystal
Back
Range for medium energy collimator
Front
200-400keV
Back
Criteria for "high radiation area"
Front
More than 1msV in 1 hour at 30cm
Back
Things that increase sensitivity of gamma camera collimator
Front
Thinner, shorter septae. Larger hole diameter
Back
How often does energy window need to be tested?
Front
Daily / before each different test using a different tracer.
Back
What happens to image from pinhole collimator
Front
Inverted and magnified
Back
Major spill of I131
Front
1mCi
Back
How often is intrinsic flood done?
Front
Weekly
Back
Where should a ring badge be worn?
Front
On the dominant hand, index finger, label facing inward toward source, under a glove
Back
Geiger counter dead time
Front
If you over load it (>100mR/h) it stops working until it dissipates.
Back
Number of disintegrations per Curie
Front
3.7 x 10^10
Back
Effective half life
Front
1 / effective = 1/physical + 1/biological
Back
How often is dose calibrator consistency checked?
Front
Daily
Back
Recommended counts for extrinsic and intrinsic flood
Front
5-10 million
Back
How often is extrinsic flood done?
Front
Once daily. This tests the collimators and the crystals
Back
Sensitivity/resolution of thick crystal
Front
Better sensitivity (chance of catching the gamma particle), worse spatial resolution (light produced moves before getting to PMT
Back
Problem with sodium iodine well counter
Front
Easily overwhelmed (if exceeds 5k counts/sec). Good for in vitro samples and wipe tests
Back
What is an isometric transition?
Front
Process of rearranging of electrons coupled to electron capture that causes gamma emission, good for imaging
Back
How often is dose calibrator accuracy checked?
Front
Annually
Back
Range for high energy collimator
Front
>400keV
Back
NRC CFR part 35
Front
Medical use of by-product material (human use of radioisotopes)
Back
Workhorse nucs collimator
Front
Parallel hole collimator
Back
Star artifact
Front
Caused by septal penetration of hexagonally oriented collimator holes by very high energy focal source (think post-therapy I131)
Back
Extrinsic flood
Front
WITH a collimator
Back
How long do you keep radioactive material?
Front
10 half lives
Back
Amount of nonuniformity that is allowable
Front
2-5% (1% in SPECT)
Back
Intrinsic flood
Front
WITHOUT a collimator
Back
How is sensitivity affected by distance for parallel hole collimator?
Front
NO change! Even though farther distance reduces counts by inverse square, it allows for greater field of view so no net change in counts.
Back
Relationship between sensitivity and spatial resolution of gamma camera collimator
Front
Inversely related
Back
What happens to image from converging hole collimator
Front
Magnifies WITHOUT inverting
Back
Nucs QC stuff that has to be done daily
Front
Extrinsic flood and energy window
Back
Mechanism of isotope production that is "carrier free"
Front
Cyclotron bombardment with charged particles -> transmutation
Back
Annual dose limit to the general public
Front
100mrem / 1mSv
Back
Impact of matrix size in gamma camera on acquisition time
Front
Larger matrix = longer acquisition time and reduced counts per pixel (worse SNR)
Back
Function of pulse height analyzer
Front
To discard background stuff that has energy too high/low to correspond to desired tracer
Back
How often does center of linearity need to be tested?
Front
Weekly
Back
How is gamma camera linearity and spatial resolution tested?
Front
By placing a lead bar phantom with parallel lines between a cobalt source and the gamma camera. Linearity is bad if the bars look wavy. Resolution is defined by ability to differentiate distinct bars
Back
How often does gamma camera image linearity and spatial resolution need to be tested?
Front
Weekly
Back
How often is dose calibrator linearity checked?
Front
Quarterly
Back
Nucs QC stuff that has to be done weekly
Front
Intrinsic flood, linearity and resolution, and center of rotation
Back
Major spill of In-111, Ga-67
Front
>10mCi
Back
How often is dose calibrator geometry checked
Front
Installation and anytime it's moved
Back
Major spill of Tc, Tl
Front
>100mCi
Back
Range for "low energy" collimator
Front
1-200keV
Back
NRC CFR part 20
Front
Radiation protection
Back
What does flood test for?
Front
Uniformity
Back
Section 6
(50 cards)
Thick dampening block (low q)
Front
"Thud" - more dampening, higher bandwidth, low Q, short spatial pulse length for better axial resolution
Back
Annual occupational lens dose limit
Front
20mSv / 2rem
Back
Situation in which there is a ton of reflection
Front
High impedance difference between adjacent tissues
Back
Total dose to embryo allowed in 9 month pregnancy
Front
5mSv / 500mrem
Back
Beam width artifact
Front
Signal from far zone falsely localizing into area of interest. Classically shown in bladder
Back
How soon do you have to report a "medical event"
Front
Call the doctor, patient, and NRC within 24 hours. Write them a letter within 15 days
Back
Side lobe artifact
Front
Artifact that happens when stuff from side lobes gets registered like it came from the main beam. Classic example = "pseudosludge" in the gallbladder or bladder
Back
Echogenicity vs scatter amplitude
Front
Hyperechoic = high scatter amplitude
Hypoechoic = lower scatter amplitude
Back
Reverberation artifact
Front
Due to reflections between two parallel highly reflective surfaces. Looks like multiple equidistantly spaced linear reflections
Back
How often do you do a "blank scan" for PET?
Front
Daily. This is the PET equivalent of uniformity
Back
Loss of how many dB corresponds with 50% loss in signal intensity?
Front
-3dB = 50% loss in signal intensity
Back
Rarefaction
Front
The relaxed (low pressure) part of the sound wave (vs compression)
Back
Type of crystal used in PET
Front
BGO, LSO, or LYSO
Back
How to tell uncorrected from attenuation correction
Front
1. Skin is hot on uncorrected
2. Lung is hot on uncorrected
Back
Type of crystal used in planar imaging
Front
NaI
Back
Truncation artifact in PET
Front
When a fat dude cant fit in a CT and part of him gets left out of the image, but he can fit in the PET, and the attenuation correction is all jacked up as a result
Back
Radiochemical purity
Front
Testing for free Tc. Use thin layer chromatography.
Back
Fasting duration prior to PET
Front
At least 4 hours, longer to minimize cardiac activity
Back
Absorption in ultrasound
Front
Sound energy gets turned into heat. This increases with frequency.
Back
Thickness of crystal ~ frequency
Front
Thick crystal - low frequency
Thin crystal - high frequency
Back
Definition of "half value thickness" in ultrasound
Front
Thickness of tissue that causes a reduction of ultrasound intensity by 3dB
Back
Criteria for "very high radiation area"
Front
More than 5 gray in 1 hour at 1m
Back
Distance an F18 positron travels in tissue
Front
~1mm
Back
Criteria for a "medical event"
Front
Wrong dose, wrong patient, wrong site (has to be off by over 20%) AND has to cause harm to the patient (whoe body dose >5rem /50mSv or single organ dose >50rem)
Back
What is impedance?
Front
Degree of "stiffness" of a tissue. Product of the velocity of sound in the medium and the density of the medium
Back
Narrow receive bandwidth results in...
Front
Better SNR, worse type 1 chemical shift, worse spatial resolution
Back
Unit of impedance
Front
Rayl
Back
Limit of how much the dose can be off from what you order
Front
20% via the NRC, 10% in some agreement states
Back
Steeper (bigger) receive bandwidth results in...
Front
Poorer SNR, better spatial resolution, better suppression of type 1 chemical shift
Back
Type of transducer that gets more side lobe artifact
Front
Linear array transducers
Back
Half life of Mo-99
Front
67 hours
Back
Minimum distance between objects in order to be resolved on axial resolution
Front
1/2 the spatial pulse length. This is smaller (better axial res) on low q / heavy damping
Back
When testing radionuclide purity, what do you assay for first?
Front
Mo first, to prevent issues with residual charge
Back
SUV estimation in fat people
Front
You'll overestimate SUV
Back
Thin damping (high Q)
Front
"Ding" - less dampening, narrow bandwidth, long spatial pulse length. good for doppler
Back
Chemical purity
Front
Testing for Al breakthrough. Test is with pH paper. Allowed amount is <10 microgram Al per 1mL
Back
Speed that US machine assumes sound travels at
Front
1540m/s
Back
What changes when ultrasound goes through different mediums? Wavelength or frequency?
Front
Speed of sound diff in diff media. As speed changes, frequency stays the same so wavelength has to change.
Back
Bucket setup imbalance
Front
Dark block rotating on sinogram
Back
How does frequency relate to scatter?
Front
Higher frequency = smaller wavelength = surfaces less smooth = more scatter
Back
Elevation resolution
Front
Resolution in the plane orthogonal to lateral resolution. Dependent on transducer element height
Back
Annual occupational total equivalent organ dose
Front
500mSv / 50rem. Same as extremity (hand) dose
Back
How often do you do a normalization scan for PET?
Front
Monthly. You scan a point source in the FOV
Back
Distribution of free Tc
Front
Salivary, thyroid, stomach
Back
What does a change of +/- 10dB have on the power
Front
+10dB = 10x power, -10db = 1/10th power
Back
Causes of free Tc
Front
Not enough tin, air in the syringe
Back
Radionuclide purity
Front
Testing for Mo breakthrough. You want less than 0.15MICROCi Mo per 1milCi of Tc - at the time of ADMINISTRATION
Back
Annual occupational whole body dose limit
Front
50mSv / 5rem
Back
Relationship between transducer frequency and near field length
Front
Higher frequency, longer near field
Back
How does axial resolution change with depth?
Front
It doesnt. Only lateral resolution does
Back
Section 7
(14 cards)
Annual MR scanner QC is done by the...
Front
Physicist
Back
Weekly MR scanner QC is done by the..
Front
Tech
Back
Crosstalk artifact
Front
Excitation into an adjacent slice, fix by reducing overlap (having a gap)
Back
Black blood cardiac MRI
Front
Double inversion recovery spin echo sequence
Back
FDA limit for specific absorption rate (SAR)
Front
4W/kg over 15 minutes
Back
Situation where neurostimulation occurs in MRI
Front
High gradient switching
Back
Dielectric effect is worse with...
Front
A stronger magnet
Back
Type of sequence in breast MRI for implant rupture
Front
Fat and water saturated (only silicone will be bright)
Back
Bright blood cardiac MRI
Front
Gradient sequences. SSFP is closer to a T2
Back
Delayed gad image sequence type cardiac MRI
Front
Inversion recovery. Null myocardium.
Back
Components of calculating SAR
Front
Magnet strength (squared)
Flip angle (squared)
Duty cycle (inversely related to TR and linear)
Back
Maximum MRI noise allowed by FDA
Front
140dB
Back
Eddy currents
Front
looks like distortion, most severe with DWI sequences
Back
Super long inversion time on cardiac imaging, to the point where blood is darker than myocardium.